Mettler toledo scale error codes

15 series · 151 error codes

Error messages on Mettler Toledo balances vary by device series. Look up stability, calibration, pan, capacity and memory warnings under the correct series.

NO STABILITY Unable to obtain a stable weighing value ME / NewClassic · The balance cannot reach a stable weighing value. WRONG ADJUSTMENT WEIGHT Wrong calibration weight ME / NewClassic · The wrong weight was used during calibration, or the required weight is not on the weighing pan. REFERENCE TOO SMALL Reference weight too low ME / NewClassic · The reference weight used for piece counting is insufficient. EEPROM ERROR EEPROM memory error ME / NewClassic · An error was detected in the permanent memory of the device. WRONG CELL DATA Weighing cell data error ME / NewClassic · The data belonging to the weighing cell is invalid or faulty. NO STANDARD ADJUSTMENT Standard adjustment data not found ME / NewClassic · NO STANDARD ADJUSTMENT indicates that the standard adjustment (calibration) data was not found on the Mettler Toledo ME / NewClassic. PROGRAM MEMORY DEFECT Program memory error ME / NewClassic · PROGRAM MEMORY DEFECT indicates a program memory error on the Mettler Toledo ME / NewClassic. TEMP SENSOR DEFECT Temperature sensor error ME / NewClassic · TEMP SENSOR DEFECT indicates a temperature sensor error on the Mettler Toledo ME / NewClassic. WRONG LOAD CELL BRAND Incompatible load cell / weighing cell ME / NewClassic · WRONG LOAD CELL BRAND indicates an incompatible load cell / weighing cell on the Mettler Toledo ME / NewClassic. WRONG TYPE DATA SET Device type data set error ME / NewClassic · WRONG TYPE DATA SET indicates a device type data set error on the Mettler Toledo ME / NewClassic. OVERLOAD Maximum capacity exceeded ME / NewClassic · The load on the weighing pan exceeds the permitted weighing capacity of the balance. UNDERLOAD Underload / weighing pan warning ME / NewClassic · UNDERLOAD indicates an underload / weighing pan warning on the Mettler Toledo ME / NewClassic. BELOW INITIAL ZERO RANGE Below ınitial zero range ME / NewClassic · BELOW INITIAL ZERO RANGE indicates a below initial zero range condition on the Mettler Toledo ME / NewClassic. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Backup battery / date and time memory warning ME / NewClassic · The backup power source that preserves the date and time may be depleted. ABOVE INITIAL ZERO RANGE Above ınitial zero range ME / NewClassic · ABOVE INITIAL ZERO RANGE indicates an above initial zero range condition on the Mettler Toledo ME / NewClassic. NO STABILITY Unable to obtain a stable weighing value MS-S · The balance cannot reach a stable weighing value. WRONG ADJUSTMENT WEIGHT Wrong calibration weight MS-S · The wrong weight was used during calibration, or the required weight is not on the weighing pan. REFERENCE TOO SMALL Reference weight too low MS-S · The reference weight used for piece counting is insufficient. EEPROM ERROR EEPROM memory error MS-S · An error was detected in the permanent memory of the device. WRONG CELL DATA Weighing cell data error MS-S · The data belonging to the weighing cell is invalid or faulty. NO STANDARD ADJUSTMENT Standard adjustment data not found MS-S · NO STANDARD ADJUSTMENT indicates that the standard adjustment (calibration) data was not found on the Mettler Toledo MS-S. PROGRAM MEMORY DEFECT Program memory error MS-S · PROGRAM MEMORY DEFECT indicates a program memory error on the Mettler Toledo MS-S. TEMP SENSOR DEFECT Temperature sensor error MS-S · TEMP SENSOR DEFECT indicates a temperature sensor error on the Mettler Toledo MS-S. WRONG LOAD CELL BRAND Incompatible load cell / weighing cell MS-S · WRONG LOAD CELL BRAND indicates an incompatible load cell / weighing cell on the Mettler Toledo MS-S. WRONG TYPE DATA SET Device type data set error MS-S · WRONG TYPE DATA SET indicates a device type data set error on the Mettler Toledo MS-S. OVERLOAD Maximum capacity exceeded MS-S · The load on the weighing pan exceeds the permitted weighing capacity of the balance. UNDERLOAD Underload / weighing pan warning MS-S · UNDERLOAD indicates an underload / weighing pan warning on the Mettler Toledo MS-S. BELOW INITIAL ZERO RANGE Below ınitial zero range MS-S · BELOW INITIAL ZERO RANGE indicates a below initial zero range condition on the Mettler Toledo MS-S. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Backup battery / date and time memory warning MS-S · The backup power source that preserves the date and time may be depleted. INITIAL ZERO RANGE EXCEEDED Initial zero range exceeded MS-S · INITIAL ZERO RANGE EXCEEDED indicates that the initial zero range was exceeded on the Mettler Toledo MS-S. MEM FULL Memory full MS-S · The record capacity of the application is full. FACTOR OUT OF RANGE Factor out of range MS-S · FACTOR OUT OF RANGE indicates that the factor is outside the permitted range on the Mettler Toledo MS-S. STEP OUT OF RANGE Step value out of range MS-S · STEP OUT OF RANGE indicates that the step value is outside the permitted range on the Mettler Toledo MS-S. OUT OF RANGE Entered value out of range MS-S · OUT OF RANGE indicates that the entered value is outside the permitted range on the Mettler Toledo MS-S. NO STABILITY Unable to obtain a stable weighing value MS-L · The balance cannot reach a stable weighing value. WRONG ADJUSTMENT WEIGHT Wrong calibration weight MS-L · The wrong weight was used during calibration, or the required weight is not on the weighing pan. REFERENCE TOO SMALL Reference weight too low MS-L · The reference weight used for piece counting is insufficient. EEPROM ERROR EEPROM memory error MS-L · An error was detected in the permanent memory of the device. WRONG CELL DATA Weighing cell data error MS-L · The data belonging to the weighing cell is invalid or faulty. NO STANDARD ADJUSTMENT Standard adjustment data not found MS-L · NO STANDARD ADJUSTMENT indicates that the standard adjustment (calibration) data was not found on the Mettler Toledo MS-L. PROGRAM MEMORY DEFECT Program memory error MS-L · PROGRAM MEMORY DEFECT indicates a program memory error on the Mettler Toledo MS-L. TEMP SENSOR DEFECT Temperature sensor error MS-L · TEMP SENSOR DEFECT indicates a temperature sensor error on the Mettler Toledo MS-L. WRONG LOAD CELL BRAND Incompatible load cell / weighing cell MS-L · WRONG LOAD CELL BRAND indicates an incompatible load cell / weighing cell on the Mettler Toledo MS-L. WRONG TYPE DATA SET Device type data set error MS-L · WRONG TYPE DATA SET indicates a device type data set error on the Mettler Toledo MS-L. OVERLOAD Maximum capacity exceeded MS-L · The load on the weighing pan exceeds the permitted weighing capacity of the balance. UNDERLOAD Underload / weighing pan warning MS-L · UNDERLOAD indicates an underload / weighing pan warning on the Mettler Toledo MS-L. BELOW INITIAL ZERO RANGE Below ınitial zero range MS-L · BELOW INITIAL ZERO RANGE indicates a below initial zero range condition on the Mettler Toledo MS-L. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Backup battery / date and time memory warning MS-L · The backup power source that preserves the date and time may be depleted. INITIAL ZERO RANGE EXCEEDED Initial zero range exceeded MS-L · INITIAL ZERO RANGE EXCEEDED indicates that the initial zero range was exceeded on the Mettler Toledo MS-L. MEM FULL Memory full MS-L · The record capacity of the application is full. FACTOR OUT OF RANGE Factor out of range MS-L · FACTOR OUT OF RANGE indicates that the factor is outside the permitted range on the Mettler Toledo MS-L. STEP OUT OF RANGE Step value out of range MS-L · STEP OUT OF RANGE indicates that the step value is outside the permitted range on the Mettler Toledo MS-L. OUT OF RANGE Entered value out of range MS-L · OUT OF RANGE indicates that the entered value is outside the permitted range on the Mettler Toledo MS-L. NO STABILITY Unable to obtain a stable weighing value MS Semi-Micro · The balance cannot reach a stable weighing value. WRONG ADJUSTMENT WEIGHT Wrong calibration weight MS Semi-Micro · The wrong weight was used during calibration, or the required weight is not on the weighing pan. REFERENCE TOO SMALL Reference weight too low MS Semi-Micro · The reference weight used for piece counting is insufficient. EEPROM ERROR EEPROM memory error MS Semi-Micro · An error was detected in the permanent memory of the device. WRONG CELL DATA Weighing cell data error MS Semi-Micro · The data belonging to the weighing cell is invalid or faulty. NO STANDARD ADJUSTMENT Standard adjustment data not found MS Semi-Micro · NO STANDARD ADJUSTMENT indicates that the standard adjustment (calibration) data was not found on the Mettler Toledo MS Semi-Micro. PROGRAM MEMORY DEFECT Program memory error MS Semi-Micro · PROGRAM MEMORY DEFECT indicates a program memory error on the Mettler Toledo MS Semi-Micro. TEMP SENSOR DEFECT Temperature sensor error MS Semi-Micro · TEMP SENSOR DEFECT indicates a temperature sensor error on the Mettler Toledo MS Semi-Micro. WRONG LOAD CELL BRAND Incompatible load cell / weighing cell MS Semi-Micro · WRONG LOAD CELL BRAND indicates an incompatible load cell / weighing cell on the Mettler Toledo MS Semi-Micro. WRONG TYPE DATA SET Device type data set error MS Semi-Micro · WRONG TYPE DATA SET indicates a device type data set error on the Mettler Toledo MS Semi-Micro. OVERLOAD Maximum capacity exceeded MS Semi-Micro · The load on the weighing pan exceeds the permitted weighing capacity of the balance. UNDERLOAD Underload / weighing pan warning MS Semi-Micro · UNDERLOAD indicates an underload / weighing pan warning on the Mettler Toledo MS Semi-Micro. BELOW INITIAL ZERO RANGE Below ınitial zero range MS Semi-Micro · BELOW INITIAL ZERO RANGE indicates a below initial zero range condition on the Mettler Toledo MS Semi-Micro. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Backup battery / date and time memory warning MS Semi-Micro · The backup power source that preserves the date and time may be depleted. INITIAL ZERO RANGE EXCEEDED Initial zero range exceeded MS Semi-Micro · INITIAL ZERO RANGE EXCEEDED indicates that the initial zero range was exceeded on the Mettler Toledo MS Semi-Micro. MEM FULL Memory full MS Semi-Micro · The record capacity of the application is full. FACTOR OUT OF RANGE Factor out of range MS Semi-Micro · FACTOR OUT OF RANGE indicates that the factor is outside the permitted range on the Mettler Toledo MS Semi-Micro. STEP OUT OF RANGE Step value out of range MS Semi-Micro · STEP OUT OF RANGE indicates that the step value is outside the permitted range on the Mettler Toledo MS Semi-Micro. OUT OF RANGE Entered value out of range MS Semi-Micro · OUT OF RANGE indicates that the entered value is outside the permitted range on the Mettler Toledo MS Semi-Micro. Balance reset failed Balance restart (reset) failed XPR Precision Balances and Comparators · There may be a communication problem between system components. The system has no valid date and time set No valid date and time ınformation XPR Precision Balances and Comparators · The internal battery level may be low. Weight cannot be determined Weight cannot be determined XPR Precision Balances and Comparators · Weight cannot be determined indicates that the weight cannot be determined on the Mettler Toledo XPR Precision Balances and Comparators. Cannot start adjustment Calibration cannot be started XPR Precision Balances and Comparators · The initial zero condition may not have been met. Preventive performance optimization Performance optimization required XPR Precision Balances and Comparators · The system memory may be heavily used. Balance reset failed Balance restart (reset) failed XSR Analytical · There may be a communication problem between system components. The system has no valid date and time set No valid date and time ınformation XSR Analytical · The internal battery level may be low. Weight cannot be determined Weight cannot be determined XSR Analytical · Weight cannot be determined indicates that the weight cannot be determined on the Mettler Toledo XSR Analytical. Cannot start adjustment Calibration cannot be started XSR Analytical · The initial zero condition may not have been met. Preventive performance optimization Performance optimization required XSR Analytical · The system memory may be heavily used. Error 1 Weighing stability not achieved HR73 · A stable weighing result could not be obtained during taring or calibration. Error 2 Wrong calibration weight HR73 · The calibration weight is missing or unsuitable. Error 3 Sample weight out of range HR73 · The sample weight is below 100 mg or outside the active weight limits. Error 4 Tare missing HR73 · The sample container has not been tared. Error 5 Invalid entry HR73 · The entered value is outside the permitted range or a method with the same name already exists. Error 6 Function not enabled in menu HR73 · The requested function cannot be used because of the menu settings. Error 7 Timer not enabled HR73 · The selected switch-off criterion does not allow a time entry. Error 8 Temperature adjustment entry timeout HR73 · The required value was not entered within the allowed time during temperature calibration. Error 9 Function error HR73 · The device detected an unidentified operating state. RAM LOST Memory backup data lost HR73 · The internal battery is depleted or a problem has occurred in the device memory. UNDERLOAD DISPLAY Underload range HR73 · The weighing value is below the lower operating range of the device. OVERLOAD DISPLAY Overload HR73 · The sample weight exceeds the capacity of the device. Error 1 Weighing stability not achieved HG53 · A stable weighing result could not be obtained during taring or calibration. Error 2 Wrong calibration weight HG53 · The calibration weight is missing or unsuitable. Error 3 Sample weight out of range HG53 · The sample weight is below 100 mg or outside the active weight limits. Error 4 Tare missing HG53 · The sample container has not been tared. Error 5 Invalid entry HG53 · The entered value is outside the permitted range or a method with the same name already exists. Error 6 Function not enabled in menu HG53 · The requested function cannot be used because of the menu settings. Error 7 Timer not enabled HG53 · The selected switch-off criterion does not allow a time entry. Error 8 Temperature adjustment entry timeout HG53 · The required value was not entered within the allowed time during temperature calibration. Error 9 Function error HG53 · The device detected an unidentified operating state. RAM LOST Memory backup data lost HG53 · The internal battery is depleted or a problem has occurred in the device memory. UNDERLOAD DISPLAY Underload range HG53 · The weighing value is below the lower operating range of the device. OVERLOAD DISPLAY Overload HG53 · The sample weight exceeds the capacity of the device. MEM.FULL Memory full ME / NewClassic · The application's record memory is full. The new value cannot be saved. FACTOR OUT OF RANGE Factor out of range ME / NewClassic · The coefficient entered in the multiplier application is not in the valid range. STEP OUT OF RANGE Step value out of range ME / NewClassic · The display step entered is not in the range allowed for the coefficient and resolution. OUT OF RANGE Sample weight outside allowed range ME / NewClassic · In the statistics application, the sample placed is outside the accepted weight range. Pairing error. Terminal pairing error XPR Precision Balances and Comparators · The connection between the terminal and the weighing platform cannot be established. ‾‾‾‾‾‾‾ (üst kesik çizgiler) Overload indicator IND231 / IND236 · Dashes at the top of the display indicate that the load is more than 9 divisions above capacity. _______ (alt kesik çizgiler) Below zero indicator IND231 / IND236 · Dashes at the bottom of the display indicate that the weight has fallen more than 20 divisions below zero. ¯no¯ / _no_ Outside zero range IND231 / IND236 · The ZERO command was not applied because the weight is outside the zeroing range. --no-- Invalid entry or key IND231 / IND236 · Indicates that an incorrect value was entered or a key that is not valid at that moment was pressed. Err 3 EEPROM verification error IND231 / IND236 · The terminal detected an error while verifying the data in the EEPROM memory. Err 35 Calibration failed IND231 / IND236 · The operation timed out because the weight did not stabilize within 30 seconds during calibration. Err 4 Too few samples IND231 / IND236 · The number of samples taken as reference in the counting application is insufficient. Err 6 EEPROM read/write error IND231 / IND236 · The terminal cannot write data to or read data from the EEPROM memory. Err 10 20d below gross zero IND231 / IND236 · With setting F1.4.2 at 20dZero, the weight fell more than 20 divisions below gross zero. Err 11 Access to setup menu blocked IND231 / IND236 · Entry to the setup menu was requested but access is locked. not CAL Scale not calibrated IND231 / IND236 · The terminal could not find a valid calibration record. no dAtA Remote display not receiving data IND231 / IND236 · While the terminal is used as a remote display, it cannot receive data from the configured port. Err 0001 Metrology switch off IND246 · An approval region is selected but the W&M (metrology) switch SW1-1 is off. Err 0002 Setting cannot be changed while approved IND246 · This setup parameter cannot be changed while the terminal is in approved status. Err 0003 Division value will be reset IND246 · Reports that the increment (division) value will be reset because the capacity was changed. Err 0004 Range 2 capacity invalid IND246 · In multi-range mode, the range 2 capacity must be greater than range 1. Err 0005 Capacity limit exceeded IND246 · The entered capacity value exceeds the permitted limit. Err 0006 Calibration weight out of range IND246 · The test weight entered for calibration is outside the permitted range. Err 0007 GEO value out of limits IND246 · The entered GEO (geographic adjustment) value is outside the permitted range. Err 0008 Invalid capacity value IND246 · The entered capacity value is not accepted by the terminal. Err 0009 Range division ratio incorrect IND246 · In multi-range mode, the ratio of the range 2 division to the range 1 division must be less than 50. E E E Zero could not be captured at startup IND246 · With startup zeroing enabled, the weight on the platform is outside the startup zeroing range. exceeds zero range (+) Above zero range IND246 · The ZERO key was pressed but the weight is above the zeroing range. illegal scale mode Zeroing not possible in net mode IND246 · The ZERO key was pressed while the terminal was in net (tared) mode. motion Command not completed due to motion IND246 · The zero, tare or print command could not be completed because motion was detected on the scale. no demand output Print connection not defined IND246 · Printing was requested but no port has a demand output defined. scale not ready Not ready to print IND246 · Printing was requested but the print lock or the auto print reset condition was not met. scale over capacity Print attempted over capacity IND246 · Printing was requested while the scale was over capacity. scale under zero Print attempted below zero IND246 · Printing was requested while the weight was below zero (display blank). tare not enabled Tare disabled IND246 · The TARE key was pressed but the tare function is turned off in the settings. tare too small Tare value too small IND246 · The TARE key was pressed when the weight was less than one division. tare under zero Tare below zero IND246 · The TARE key was pressed while the weight was below zero. tare exceeds limit Preset tare greater than capacity IND246 · The entered preset tare value is greater than the scale capacity. keypad tare disabled Keypad tare off IND246 · Preset tare entry was attempted but keypad tare is turned off in the settings. PDX CELL xx NO RESPONSE POWERCELL cell not responding IND246 · In the POWERCELL version, load cell number xx is not responding to the terminal. PDX CELL xx NOT FOUND POWERCELL cell not found IND246 · Load cell number xx could not be detected on the POWERCELL network. EEE Zero could not be captured at startup IND570 · With startup zeroing enabled, the weight is outside the zero capture range and the terminal shows EEE (error log A00024). Scale blanked over capacity Capacity exceeded - display blank IND570 · The display was blanked because the weight passed the set capacity overload limit. Scale under zero Weight below zero IND570 · The weight fell below the below-zero blanking limit. Zero failed - Range Zeroing failed - range IND570 · The zero command was not applied because the weight is outside the permitted zeroing range. Tare Failed–Motion Tare could not be taken due to motion IND570 · The TARE command was issued but the scale did not stabilize within the timeout period. Tare Failed–Below Zero Tare could not be taken below zero IND570 · Tare was attempted while the gross weight was zero or below. Low excitation voltage Low excitation voltage IND570 · The excitation voltage to the load cells fell below 9.0 V and the service icon lights. Excitation voltage below acceptable limit Excitation voltage below acceptable limit IND570 · The load cell excitation voltage fell below 5.0 V and a pop-up error message was shown. Potential analog section saturation. Recalibrate with new capacity. Analog input saturation IND570 · It was detected that the analog measurement section may reach saturation. Recalibration with the new capacity is recommended. Battery voltage extremely low Internal battery voltage too low IND570 · The voltage of the internal battery that preserves setup data is critically low. Battery voltage low Internal battery voltage low IND570 · The internal (RAM/BRAM) battery voltage is at a low level. Internal Flash error Internal flash memory error IND570 · An error was detected in the terminal's internal Flash memory. Node * no response POWERCELL node not responding IND570 · Node number * (load cell) on the POWERCELL network is not responding. Node * not found POWERCELL node not found IND570 · Node number * on the POWERCELL network cannot be found by the terminal. Node * negative out of range POWERCELL node below negative range IND570 · Cell number * on the POWERCELL network is reporting a value below the range (negative). Connect IDNet base Idnet platform not connected IND570 · The terminal does not detect an IDNet weighing platform. Connect SICSpro scale Sıcspro platform not connected IND570 · The terminal does not detect a SICSpro weighing platform. No digital scale interface No digital scale interface IND570 · The terminal does not detect a digital scale interface connected to the main board. Low supply voltage Low supply voltage IND570 · The terminal's supply voltage was detected as low. Platform battery low Platform battery low IND570 · The battery life of the wireless (Cable-Free) floor scale platform has dropped to about one month. Bluetooth error – IND570 Terminal bluetooth module error IND570 · The Bluetooth module installed in the IND570 is not communicating. Bluetooth error- scale Platform bluetooth module error IND570 · The Bluetooth module in the scale base is not communicating. Bluetooth OK - No scale data Bluetooth connected, no scale data IND570 · The Bluetooth connection is active but the platform is not sending weight data. Calibration expired. Scale disabled. Run Calibration test. Calibration expired IND570 · The calibration validity period has expired and the scale was disabled because of the 'Alarm & Disable' setting. Calibration Failure Calibration failed IND570 · Zero/span capture or CalFree calibration failed. NO STABILITY No stability ML / NewClassic · The balance cannot get a stable weight reading. WRONG ADJUSTMENT WEIGHT Wrong adjustment weight ML / NewClassic · The required adjustment weight is not on the pan. REFERENCE TOO SMALL Reference weight too low ML / NewClassic · The reference for piece counting is too light. EEPROM ERROR - PLEASE CONTACT CUSTOMER SERVICE EEPROM error ML / NewClassic · The memory that stores the balance settings has an error. WRONG CELL DATA - PLEASE CONTACT CUSTOMER SERVICE Wrong cell data ML / NewClassic · The weighing cell data is read incorrectly. NO STANDARD ADJUSTMENT - PLEASE CONTACT CUSTOMER SERVICE No standard adjustment ML / NewClassic · The basic factory adjustment data is missing. PROGRAM MEMORY DEFECT - PLEASE CONTACT CUSTOMER SERVICE Program memory defect ML / NewClassic · The memory holding the balance firmware is faulty. TEMP SENSOR DEFECT - PLEASE CONTACT CUSTOMER SERVICE Temperature sensor defect ML / NewClassic · The internal temperature sensor does not work. WRONG LOAD CELL BRAND - PLEASE CONTACT CUSTOMER SERVICE Wrong load cell brand ML / NewClassic · The fitted weighing cell does not match the electronics. WRONG TYPE DATA SET - PLEASE CONTACT CUSTOMER SERVICE Wrong type data set ML / NewClassic · The model data of the balance is wrong. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Battery backup lost ML / NewClassic · The internal battery that keeps the date and time is empty. Overload Capacity exceeded ML / NewClassic · The load on the pan exceeds the capacity. Underload Underload ML / NewClassic · The balance reads a load lower than expected. ABOVE INITIAL ZERO RANGE Above ınitial zero range ML / NewClassic · The load at power-on is above the zero limit. BELOW INITIAL ZERO RANGE Below ınitial zero range ML / NewClassic · The reading at power-on is below the zero limit. MEM.FULL Memory full ML / NewClassic · The application memory is full. FACTOR OUT OF RANGE Factor out of range ML / NewClassic · The entered factor is outside the allowed range. STEP OUT OF RANGE Step out of range ML / NewClassic · The entered step is outside the allowed range. OUT OF RANGE Out of range ML / NewClassic · The sample weight is outside the range accepted by the application. NO STABILITY No stability JE Jewelry · The balance cannot get a stable weight reading. WRONG ADJUSTMENT WEIGHT Wrong adjustment weight JE Jewelry · The required adjustment weight is not on the pan. REFERENCE TOO SMALL Reference weight too low JE Jewelry · The reference for piece counting is too light. EEPROM ERROR - PLEASE CONTACT CUSTOMER SERVICE EEPROM error JE Jewelry · The memory that stores the balance settings has an error. WRONG CELL DATA - PLEASE CONTACT CUSTOMER SERVICE Wrong cell data JE Jewelry · The weighing cell data is read incorrectly. NO STANDARD ADJUSTMENT - PLEASE CONTACT CUSTOMER SERVICE No standard adjustment JE Jewelry · The basic factory adjustment data is missing. PROGRAM MEMORY DEFECT - PLEASE CONTACT CUSTOMER SERVICE Program memory defect JE Jewelry · The memory holding the balance firmware is faulty. TEMP SENSOR DEFECT - PLEASE CONTACT CUSTOMER SERVICE Temperature sensor defect JE Jewelry · The internal temperature sensor does not work. WRONG LOAD CELL BRAND - PLEASE CONTACT CUSTOMER SERVICE Wrong load cell brand JE Jewelry · The fitted weighing cell does not match the electronics. WRONG TYPE DATA SET - PLEASE CONTACT CUSTOMER SERVICE Wrong type data set JE Jewelry · The model data of the balance is wrong. BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Battery backup lost JE Jewelry · The internal battery that keeps the date and time is empty. Overload Capacity exceeded JE Jewelry · The load on the pan exceeds the capacity. Underload Underload JE Jewelry · The balance reads a load lower than expected. ABOVE INITIAL ZERO RANGE Above ınitial zero range JE Jewelry · The load at power-on is above the zero limit. BELOW INITIAL ZERO RANGE Below ınitial zero range JE Jewelry · The reading at power-on is below the zero limit. MEM.FULL Memory full JE Jewelry · The application memory is full. FACTOR OUT OF RANGE Factor out of range JE Jewelry · The entered factor is outside the allowed range. STEP OUT OF RANGE Step out of range JE Jewelry · The entered step is outside the allowed range. OUT OF RANGE Out of range JE Jewelry · The sample weight is outside the range accepted by the application. Overload Capacity exceeded XS Precision · The load on the pan exceeds the capacity. Underload Underload XS Precision · The balance reads a load lower than expected. Error when switching on or zeroing Error when switching on or zeroing XS Precision · Limits were exceeded at power-on or zeroing and the weight display flashes. Taring or zeroing was interrupted Taring or zeroing was ınterrupted XS Precision · No stable result was obtained during taring or zeroing. Overload Capacity exceeded XS Analytical · The load on the pan exceeds the capacity. Underload Underload XS Analytical · The balance reads a load lower than expected. Error when switching on or zeroing Error when switching on or zeroing XS Analytical · Limits were exceeded at power-on or zeroing and the weight display flashes. Taring or zeroing was interrupted Taring or zeroing was ınterrupted XS Analytical · No stable result was obtained during taring or zeroing.