Message on display Potential analog section saturation. Recalibrate with new capacity. Mettler Toledo IND570

Analog input saturation

The capacity and calibration settings can cause the load cell signal to exceed the measurement range. In this case high weights cannot be measured accurately.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Ask the responsible person whether the capacity and calibration settings were changed.
  2. Request service for recalibration.

Possible causes

  • Capacity setting not suited to the load cell
  • Incorrect calibration

Do you need service?

Redoing the capacity and analog calibration requires technical service.

Models showing this code: IND570, IND570xx