Unable to obtain a stable weighing value
During taring, zeroing or adjustment (calibration), the balance waits for the reading to settle. Because the value keeps fluctuating, the operation cannot be completed. According to the manual, the solution is to improve the ambient conditions, or if that does not help, to change the environment setting.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Close the draft shield doors and block drafts from air conditioners, fans and windows near the balance.
- Move the balance to a solid, vibration-free weighing table and level it again using the bubble level.
- Make sure nothing (cable, draft shield part, sample container) touches the weighing pan.
- If conditions cannot be improved, set the environment (ENV) setting in the menu to a level suited to less stable environments and repeat the operation.
Possible causes
- Vibration on the table or floor
- Air drafts or an open draft shield door
- Sample evaporation or static electricity
- Environment setting not suited to the current conditions
Do you need service?
If the value still does not settle after the environment has been improved and the environment setting changed, the weighing cell must be checked by service.
Models showing this code: ME, NewClassic ME
Source: Mettler Toledo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.