Wrong calibration weight
During adjustment (calibration) with an external weight, the weight the balance expects is not on the weighing pan. If the wrong weight is used or no weight is placed, the adjustment is not accepted and the previous adjustment is kept.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Read the nominal value requested on the display and prepare an adjustment weight of exactly that value.
- Empty the weighing pan and restart the adjustment.
- When prompted, place the adjustment weight in the exact center of the weighing pan and wait until the display confirms.
Possible causes
- An adjustment weight different from the one requested on the display was placed
- The adjustment weight was not placed at all, or was placed too late
- Another load besides the adjustment weight is on the weighing pan
Do you need service?
The error clears when the correct adjustment weight is placed in the center of the weighing pan. No service is required.
Models showing this code: ME, NewClassic ME
Source: Mettler Toledo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.