Code on display OVERLOAD DISPLAY Mettler Toledo HG53

Overload

The total of the sample and container on the pan exceeds the weighing capacity of the device. When the capacity is exceeded, measurement is not possible and the weighing system is strained.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Reduce the amount of sample.
  2. Use a lightweight sample container suited to the device.
  3. Check that the total weight is below the capacity of the device.
  • Do not press on the pan handler or drop loads onto it.

Possible causes

  • Too much sample added
  • Heavy or unsuitable sample container used

Do you need service?

The indication disappears when the load is brought below capacity. If it also appears with an empty pan, service is needed.

Models showing this code: HG53

Source: Mettler Toledo official user manual