Sample weight out of range
The amount of sample placed is below the minimum weight required for drying (100 mg) or outside the set weight limits. In this case moisture measurement is not started because it would not be reliable.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Confirm the message with the «Reset» key.
- Weigh at least 100 mg of sample (preferably at the target weight defined in the method).
- Restart the measurement.
Possible causes
- Sample amount less than 100 mg
- Sample outside the defined target weight limits
Do you need service?
The error clears when a suitable amount of sample is placed. No service is needed.
Models showing this code: HR73
Source: Mettler Toledo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.