Message on display Error when switching on or zeroing Mettler Toledo XS Precision

Error when switching on or zeroing

One or more limits were exceeded when the balance was switched on, woke from standby or was zeroed. The usual cause is a load on the pan at power-on.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan.
  2. Check that the weighing pan is installed correctly.
  3. Zero the balance again.

Possible causes

  • There was a load on the weighing pan at switch-on
  • Pan installed incorrectly

Do you need service?

If the message remains with an empty pan, the zero point must be checked by a technician.

Source: Mettler Toledo official user manual · Open manufacturer document