Code on display TEMP SENSOR DEFECT Mettler Toledo ME / NewClassic

Temperature sensor error

The internal temperature sensor of the balance is faulty. Weighing accuracy is affected because automatic correction for temperature changes and temperature-dependent automatic adjustment rely on this sensor.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the balance off, unplug it, and switch it back on after 30 seconds. If the message appears again, contact service.
  • Do not open the housing. The internal parts cannot be repaired by the user.
  • Note the error code and serial number and report them to technical service.

Possible causes

  • Temperature sensor failure
  • Sensor connection or electronics board problem

Do you need service?

The temperature sensor is inside the device, so it must be checked and replaced by technical service.

Models showing this code: ME, NewClassic ME

Source: Mettler Toledo official user manual