Code on display NO STANDARD ADJUSTMENT Mettler Toledo ME / NewClassic

Standard adjustment data not found

The standard adjustment (calibration) data made at the factory was not found on the balance. Without this basic adjustment, the balance cannot weigh accurately.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the balance off, unplug it, and switch it back on after 30 seconds. If the message appears again, contact service.
  • Do not open the housing. The internal parts cannot be repaired by the user.
  • Note the error code and serial number and report them to technical service.

Possible causes

  • Factory/standard adjustment data was deleted or corrupted
  • Memory or electronics board problem

Do you need service?

The standard adjustment can only be redone by service, so technical service is required.

Models showing this code: ME, NewClassic ME