Code on display Error 5 Mettler Toledo HG53

Invalid entry

The entered value is outside the accepted lower/upper limits or a method with the same name is already saved. The setting is not saved until a valid entry is made.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Confirm the message with the «Reset» key.
  2. Re-enter the value within the permitted range.
  3. If you are saving a new method, give it a different name.

Possible causes

  • Entering a value outside the allowed range
  • Reusing an existing method name

Do you need service?

The error clears when a valid value is entered. No service is needed.

Models showing this code: HG53

Source: Mettler Toledo official user manual