Message on display Weight cannot be determined Mettler Toledo XPR Precision Balances and Comparators

Weight cannot be determined

The weight value cannot be calculated because the balance is not receiving a valid weight signal from the electronics. The manual lists a data signal problem or a device error as possible causes.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Unplug the power cable, wait a few seconds and plug it in again.
  2. Open the "Device errors" section of the service menu, check whether an error is recorded, and note the error code.
  • If a device error is listed in the service menu, report the error code to technical service.

Possible causes

  • Data signal problem in the electronics
  • Device error (recorded in the service menu)

Do you need service?

If the error persists after a restart or a device error is recorded in the service menu, contact technical service with the error code.

Models showing this code: XPR Precision

Source: Mettler Toledo official user manual · Open manufacturer document