Code on display Error 2 Mettler Toledo HR73

Wrong calibration weight

The requested adjustment weight was not placed during balance adjustment (calibration) or the weight placed is wrong. The adjustment is not accepted in this case.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Confirm the message with the «Reset» key.
  2. Prepare the correct adjustment weight with the value shown on the display.
  3. Restart the adjustment and, when prompted, place the weight in the center of the pan.

Possible causes

  • Adjustment weight not placed at all
  • Adjustment weight of the wrong value used

Do you need service?

The error clears when the correct adjustment weight is used. No service is needed.

Models showing this code: HR73

Source: Mettler Toledo official user manual