Weighing stability not achieved
The operation could not be completed because the weighing value did not stabilize during taring or adjustment (calibration). You cannot continue working with the device until the error message is confirmed with the «Reset» key.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Confirm the message with the «Reset» key.
- Place the device on a solid surface free of vibration and air drafts.
- Check that the sample container and the sample are not touching the draft shield element or the pan handler.
- Repeat the tare or adjustment operation.
- Do not touch the sample chamber while the heating unit is hot.
Possible causes
- Vibration or air drafts
- Sample or sample container touching the draft shield element or the pan handler
- Device on an unstable surface
Do you need service?
If the value still does not stabilize after the environment is improved, the weighing system must be checked by service.
Models showing this code: HR73
Source: Mettler Toledo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.