Code on display Error 1 Mettler Toledo HR73

Weighing stability not achieved

The operation could not be completed because the weighing value did not stabilize during taring or adjustment (calibration). You cannot continue working with the device until the error message is confirmed with the «Reset» key.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Confirm the message with the «Reset» key.
  2. Place the device on a solid surface free of vibration and air drafts.
  3. Check that the sample container and the sample are not touching the draft shield element or the pan handler.
  4. Repeat the tare or adjustment operation.
  • Do not touch the sample chamber while the heating unit is hot.

Possible causes

  • Vibration or air drafts
  • Sample or sample container touching the draft shield element or the pan handler
  • Device on an unstable surface

Do you need service?

If the value still does not stabilize after the environment is improved, the weighing system must be checked by service.

Models showing this code: HR73

Source: Mettler Toledo official user manual