Message on display BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Mettler Toledo JE Jewelry

Battery backup lost

This battery keeps the date and time when the balance is unplugged. Because it is empty, the date and time may be lost.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Plug in the balance and leave it overnight to charge the battery.
  2. Check and correct the date and time.
  3. Check the next day whether the message comes back.

Possible causes

  • Balance unplugged for a long time
  • Battery at end of life

Do you need service?

If the date and time are still lost after charging, the internal battery must be replaced by a technician.

Models showing this code: JE703C, JE3002GE

Source: Mettler Toledo official user manual · Open manufacturer document