Message on display Preventive performance optimization Mettler Toledo XSR Analytical

Performance optimization required

The working memory (RAM) of the balance is full, so the system recommends a restart to maintain performance. To prevent data loss, the current job should be finished first.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Complete the weighing task in progress and save the results.
  2. Unplug the power cable, wait a few seconds and plug it in again.

Possible causes

  • Operating for a long time without a restart
  • Working memory (RAM) is full

Do you need service?

If the message appears again shortly after the restart, contact service for a software check.

Models showing this code: XSR Analytical

Source: Mettler Toledo official user manual · Open manufacturer document