Code on display no dAtA Mettler Toledo IND231 / IND236

Remote display not receiving data

In remote display mode, the screen shows weight data coming from the main terminal. This message appears when the data stops. The problem is usually in the port settings or wiring.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the settings of the port configured as the remote display (rCOM n).
  2. Verify that the main terminal is programmed for data output (continuous output).
  3. Check the cable and connections between the two devices.

Possible causes

  • rCOM port settings incorrect
  • Main terminal not programmed for data output
  • Broken or loose cable

Do you need service?

If no data arrives although the settings and cable are correct, the serial port hardware must be tested by service.

Models showing this code: IND231, IND236

Source: Mettler Toledo official user manual · Open manufacturer document