Code on display Connect SICSpro scale Mettler Toledo IND570

Sıcspro platform not connected

In the SICSpro version, weight data comes from the digital platform. If the platform is not detected, weighing is not possible.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform cable is correctly connected to the terminal.
  2. Turn off the terminal, disconnect the platform and wait 30 seconds, then reconnect and turn it on.

Possible causes

  • Platform cable loose
  • Cable damage
  • Platform fault

Do you need service?

If the message persists although the connection is correct, technical service is required.

Models showing this code: IND570, IND570xx

Source: Mettler Toledo official user manual · Open manufacturer document