Code on display UNDERLOAD DISPLAY Mettler Toledo HR73

Underload range

The weighing system is measuring a weight below the lowest expected value. It is usually caused by the pan handler or sample container not being seated correctly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the pan handler and the draft shield element are installed correctly.
  2. Make sure nothing is touching the pan handler.
  3. Switch the device off and on again.
  • Do not touch the sample chamber while the heating unit is hot.

Possible causes

  • Pan handler not installed or seated wrongly
  • Draft shield element installed incorrectly

Do you need service?

If the indication persists although the parts are installed correctly, the weighing system must be checked by service.

Models showing this code: HR73

Source: Mettler Toledo official user manual