Message on display Weight cannot be determined Mettler Toledo XSR Analytical

Weight cannot be determined

The weight value cannot be calculated because the balance is not receiving a valid weight signal from the electronics. The manual lists a data signal problem, a device error and the terminal-weighing unit connection as possible causes.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Unplug the power cable, wait a few seconds and plug it in again.
  2. Check the terminal cable for crushing, kinks or broken pins and make sure it is fully seated.
  3. Open the "Device errors" section of the service menu, check whether an error is recorded, and note the error code.
  • If a device error is listed in the service menu, report the error code to technical service.

Possible causes

  • Data signal problem in the electronics
  • Poor cable connection between the terminal and the weighing unit
  • Device error (recorded in the service menu)

Do you need service?

If the error persists after a restart or a device error is recorded in the service menu, contact technical service with the error code.

Models showing this code: XSR Analytical