Outside zero range
The terminal can zero only weight within the set zeroing range. If it is indicated at the top there is excess load, and if at the bottom there is a missing load. This shows that the weight is far from the calibrated zero.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Remove the load from the platform and press the ZERO key again.
- Check that all of the initial weight, such as the platform and weighing pan, is on the platform.
Possible causes
- Load left on the platform
- Initial dead load (platform, container) missing
Do you need service?
Zeroing can be done once the load is removed and the initial weight is complete. No service is needed.
Models showing this code: IND231, IND236
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.