Code on display RAM LOST Mettler Toledo HG53

Memory backup data lost

The backed-up memory content where the device stores user settings and saved drying data has been lost. Menu settings and methods may have returned to default values.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Confirm the message with the «Reset» key.
  2. Check the menu settings, date/time and drying methods, and re-enter them if needed.
  3. Check the balance and temperature adjustment before starting measurements.

Possible causes

  • Backup battery that keeps the memory alive is depleted
  • Device left without power for a long time
  • Memory problem

Do you need service?

If the message appears at every power-on, the backup battery or memory must be replaced by service.

Models showing this code: HG53

Source: Mettler Toledo official user manual