Code on display WRONG ADJUSTMENT WEIGHT Mettler Toledo MS-L

Wrong calibration weight

During adjustment (calibration) with an external weight, the weight the balance expects is not on the weighing pan. If the wrong weight is used or no weight is placed, the adjustment is not accepted and the previous adjustment is kept.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Read the nominal value requested on the display and prepare an adjustment weight of exactly that value.
  2. Empty the weighing pan and restart the adjustment.
  3. When prompted, place the adjustment weight in the exact center of the weighing pan and wait until the display confirms.

Possible causes

  • An adjustment weight different from the one requested on the display was placed
  • The adjustment weight was not placed at all, or was placed too late
  • Another load besides the adjustment weight is on the weighing pan

Do you need service?

The error clears when the correct adjustment weight is placed in the center of the weighing pan. No service is required.

Models showing this code: MS-L

Source: Mettler Toledo official user manual · Open manufacturer document