Code on display REFERENCE TOO SMALL Mettler Toledo ML / NewClassic

Reference weight too low

In piece counting the balance calculates the piece weight from the reference. If the reference is too light the result is not reliable and it is rejected.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Put more pieces on the pan for the reference.
  2. Increase the selected reference quantity.
  3. Take the reference again.

Possible causes

  • Too few pieces placed for the reference
  • Pieces being too light

Do you need service?

Counting works normally once the reference weight is increased.

Models showing this code: ML104, ML204, ML303, ML3002

Source: Mettler Toledo official user manual · Open manufacturer document