Scale not calibrated
The displayed weight is not accurate without calibration. It can appear after first installation, a reset or data loss.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Calibrate from the calibration menu (F1.3) with a suitable calibration weight.
- After calibration, check the accuracy with a known weight.
- On legal-for-trade scales, calibration must be performed only by technical service.
Possible causes
- Calibration never performed
- Reverted to factory settings
- Calibration data lost
Do you need service?
If calibration cannot be done or the message recurs, service must check the calibration and memory.
Models showing this code: IND231, IND236
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.