Code on display Err 0001 Mettler Toledo IND246

Metrology switch off

When approved use is selected, the terminal requires switch SW1-1 to be on before exiting setup. This is to protect the legal metrology settings.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. If approved use is not needed, remove the approval region in the settings.
  2. If approved use is needed, leave the operation to technical service.
  • Do not change the sealed metrology switch yourself.

Possible causes

  • Approval region selected and SW1-1 not turned on
  • Approval region selected by mistake

Do you need service?

Setting and sealing the SW1-1 switch must be done by technical service.

Models showing this code: IND246, IND246 POWERCELL

Source: Mettler Toledo official user manual · Open manufacturer document