Metrology switch off
When approved use is selected, the terminal requires switch SW1-1 to be on before exiting setup. This is to protect the legal metrology settings.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- If approved use is not needed, remove the approval region in the settings.
- If approved use is needed, leave the operation to technical service.
- Do not change the sealed metrology switch yourself.
Possible causes
- Approval region selected and SW1-1 not turned on
- Approval region selected by mistake
Do you need service?
Setting and sealing the SW1-1 switch must be done by technical service.
Models showing this code: IND246, IND246 POWERCELL
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.