Code on display Pairing error. Mettler Toledo XPR Precision Balances and Comparators

Terminal pairing error

The balance's terminal (display unit) and weighing platform cannot pair with each other. This message appears when the cable connection between them is poor. Weighing is not possible without a connection.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn off the balance, disconnect the terminal cable at both ends, and reconnect it firmly.
  2. Inspect the cable for crushing, bending, and broken pins.
  3. Turn the balance on again. If the error persists, replace the terminal cable with a new one.
  • Disconnect power to the balance before plugging or unplugging the cable.

Possible causes

  • Terminal cable damaged (crushed, bent, or broken pin)
  • Cable not fully seated

Do you need service?

If pairing is still not achieved after the cable is replaced, technical service is required for the terminal or platform electronics.

Models showing this code: XPR Precision

Source: Mettler Toledo official user manual · Open manufacturer document