Terminal pairing error
The balance's terminal (display unit) and weighing platform cannot pair with each other. This message appears when the cable connection between them is poor. Weighing is not possible without a connection.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn off the balance, disconnect the terminal cable at both ends, and reconnect it firmly.
- Inspect the cable for crushing, bending, and broken pins.
- Turn the balance on again. If the error persists, replace the terminal cable with a new one.
- Disconnect power to the balance before plugging or unplugging the cable.
Possible causes
- Terminal cable damaged (crushed, bent, or broken pin)
- Cable not fully seated
Do you need service?
If pairing is still not achieved after the cable is replaced, technical service is required for the terminal or platform electronics.
Models showing this code: XPR Precision
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.