Code on display ABOVE INITIAL ZERO RANGE Mettler Toledo ME / NewClassic

Above ınitial zero range

When the balance took its initial zero at switch-on, the measured value was above the permitted range. According to the manual, the weighing pan is not empty or the wrong (heavier) weighing pan is installed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  2. Check that the correct weighing pan for this balance model is installed.
  3. After making sure nothing touches the weighing pan, switch the balance off and on again.

Possible causes

  • A load or container was left on the weighing pan at switch-on
  • A (heavier) weighing pan that does not belong to this model is used
  • Something is touching or pressing on the weighing pan

Do you need service?

If the error persists although the weighing pan is empty and correct, a service check is required for zero point drift.

Models showing this code: ME, NewClassic ME