Code on display Underload Mettler Toledo XS Precision

Underload

The message appears if the pan is not seated correctly, cannot move freely or catches on the draft shield.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the weighing pan is installed correctly.
  2. Make sure the pan can move freely.
  3. Check that the pan does not touch the draft shield parts.

Possible causes

  • Pan missing or seated incorrectly
  • Pan touching the draft shield
  • Object caught under the pan

Do you need service?

If the message remains with the pan free and fitted correctly, the weighing cell must be checked by a technician.

Source: Mettler Toledo official user manual · Open manufacturer document