Message on display BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS Mettler Toledo MS-S

Backup battery / date and time memory warning

The backup battery that keeps the date and time when the balance is unplugged is depleted, so the date and time setting may have been lost. According to the manual, the battery charges as long as the balance stays connected to mains power.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press any key and set the date and time again.
  2. Leave the balance plugged in for a while (e.g. overnight) so the battery can charge.
  3. The next day, switch the balance off and on to check that the date and time are retained.

Possible causes

  • The balance was left unplugged for a long time
  • The backup battery is depleted or has reached the end of its life

Do you need service?

If the date and time are still lost after charging, the backup battery must be replaced by service.

Models showing this code: MS-S

Source: Mettler Toledo official user manual · Open manufacturer document