Message on display _______ (alt kesik çizgiler) Mettler Toledo IND231 / IND236

Below zero indicator

After the zero point was taken, a load may have been removed from the platform or a pan/platform part may have come loose. In this case the terminal does not show the negative value.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the platform and the top plate are in place.
  2. Empty the platform and zero it with the ZERO key.

Possible causes

  • Load on the platform when zeroing was performed
  • Platform or weighing pan dislodged
  • Load cell connection problem

Do you need service?

If zeroing with an empty platform does not solve the problem, the load cell and its connections must be checked by service.

Models showing this code: IND231, IND236

Source: Mettler Toledo official user manual · Open manufacturer document