Code on display Err 35 Mettler Toledo IND231 / IND236

Calibration failed

The new calibration values could not be saved because the calibration weight could not be read. It is shown on the display for 2 seconds and the terminal returns to its last operation.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Make sure the platform is stable and not touching anything.
  2. Reposition the calibration weight and repeat the calibration.
  3. If you wish, cancel the operation and restart when the environment has calmed down.

Possible causes

  • Vibration or air drafts
  • Calibration weight moving
  • Platform touching something

Do you need service?

If calibration still fails in a stable environment, the load cell or terminal must be inspected by service.

Models showing this code: IND231, IND236