Code on display WRONG CELL DATA Mettler Toledo MS-L

Weighing cell data error

The data stored for the weighing cell (load cell) is invalid or incompatible with the device. Without this data, the balance cannot calculate weight correctly.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the balance off, unplug it, and switch it back on after 30 seconds. If the message appears again, contact service.
  • Do not open the housing. The internal parts cannot be repaired by the user.
  • Note the error code and serial number and report them to technical service.

Possible causes

  • Corruption of the weighing cell data
  • Incompatible data after an electronics board or weighing cell replacement

Do you need service?

Reloading the weighing cell data requires service software, so technical service must be contacted.

Models showing this code: MS-L

Source: Mettler Toledo official user manual · Open manufacturer document