Code on display Err 11 Mettler Toledo IND231 / IND236

Access to setup menu blocked

In approved (legal-for-trade) use, the setup menu is protected by the metrology switch. This prevents unauthorized changes to settings.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. If you do not need to perform an operation that requires setup, return to normal weighing.
  2. If a setting change is needed, contact an authorized person or service.
  • On legal-for-trade scales, do not touch the sealed metrology switch yourself.

Possible causes

  • Metrology switch in locked position
  • Terminal in legal-for-trade mode

Do you need service?

Access to the sealed metrology switch and the re-verification afterward must be done by technical service.

Models showing this code: IND231, IND236

Source: Mettler Toledo official user manual · Open manufacturer document