Code on display UNDERLOAD Mettler Toledo MS-L

Underload / weighing pan warning

The balance is measuring a value below the lowest expected weight with the weighing pan installed, and an underload indicator appears on the display. According to the manual, this is mostly caused by the weighing pan not being seated correctly.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the weighing pan and pan support are installed and fully seated.
  2. Make sure nothing touches the weighing pan from below or around it.
  3. Switch the balance off and on again so that zeroing is performed again.

Possible causes

  • The weighing pan is not installed or is seated incorrectly
  • The pan support is missing or installed incorrectly
  • The weighing pan is rubbing against something from below or is lifted

Do you need service?

If the underload indicator remains although the weighing pan is installed correctly, contact service to have the weighing cell checked.

Models showing this code: MS-L

Source: Mettler Toledo official user manual · Open manufacturer document