Zeroing failed - range
The terminal zeroes only weight near the calibrated zero. Larger loads must be handled with tare.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty and clean the platform.
- Issue the ZERO command again.
- To subtract the container weight, use TARE instead of ZERO.
Possible causes
- Load on the platform
- Dirt/material built up on the platform
- Platform part missing
Do you need service?
If zeroing still fails on an empty, clean platform, contact service for a calibration check.
Models showing this code: IND570, IND570xx
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.