Code on display Low excitation voltage Mettler Toledo IND570

Low excitation voltage

If the excitation voltage is low, weight measurement may be inaccurate. The cause is usually a short circuit, a damp cable or a power supply problem.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. On the Terminal Status screen, check whether the excitation value is 'OK'.
  2. Check the load cell cable for visible damage.
  • Do not open the terminal cover. Electrical tests must be done only by an authorized electrician/service.

Possible causes

  • Short circuit or moisture in the load cell cable
  • Power supply fault
  • Main board fault

Do you need service?

Power supply, wiring and main board measurements must be done by service.

Models showing this code: IND570, IND570xx