Code on display E E E Mettler Toledo IND246

Zero could not be captured at startup

The terminal does not allow weighing because it could not capture the zero point at startup. It is usually caused by a load on the platform at startup.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove all load from the platform.
  2. Check that the platform and the top plate are in place.
  3. Turn off the terminal and turn it on again with an empty platform.

Possible causes

  • Load on the platform at power-on
  • Platform part missing or dislodged
  • Load cell connection problem

Do you need service?

If E E E still appears when started with an empty platform, the load cell and connections must be checked by service.

Models showing this code: IND246, IND246 POWERCELL

Source: Mettler Toledo official user manual · Open manufacturer document