Code on display PDX CELL xx NOT FOUND Mettler Toledo IND246

POWERCELL cell not found

The terminal cannot find a cell on the network that is in the configuration. The weight of this cell cannot be included.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Note the cell number shown on the display.
  2. Turn the terminal off and on again. If the message persists, contact service.
  • Do not disconnect the load cell cables yourself.

Possible causes

  • Cell or home run cable connection broken
  • Cell replaced but not addressed
  • Cell fault

Do you need service?

Cell addressing and cable/cell faults must be resolved by technical service.

Models showing this code: IND246, IND246 POWERCELL

Source: Mettler Toledo official user manual · Open manufacturer document