Maximum capacity exceeded
The weight on the weighing pan exceeds the weighing capacity of the balance. An overload indicator appears on the display instead of a value. Leaving a load above capacity on the pan can damage the weighing cell.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Reduce or remove the load on the weighing pan immediately.
- Make sure the total weight, including the tare container, is below the capacity of the balance.
- Weigh the sample in smaller portions or use a balance with a higher capacity.
- Do not place a load above capacity on the weighing pan and do not press on the pan, as the weighing cell can be damaged.
Possible causes
- The sample is heavier than the balance capacity
- The combined weight of the tare container and sample exceeds the capacity
- A load was dropped or pressed onto the weighing pan
Do you need service?
The indicator disappears when the load is brought below capacity. If it also appears on an empty pan, a service check is required.
Models showing this code: MS-L
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.