Code on display OUT OF RANGE Mettler Toledo ME / NewClassic

Sample weight outside allowed range

In the statistics application, the sample weight placed was not recorded because it is outside the accepted range. The next sample must be within approximately 70%–130% of the current mean.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Place a new sample close to the current mean (within approximately 70%–130%).
  3. Start a new statistics series for samples of different weights.

Possible causes

  • The sample differs too much from the current mean
  • The wrong sample, or more than one sample, was placed

Do you need service?

The weighing is accepted when a sample within the proper range is placed. No service is required.

Models showing this code: ME, NewClassic ME

Source: Mettler Toledo official user manual