Code on display NO STABILITY Mettler Toledo ML / NewClassic

No stability

The reading keeps changing, so the balance cannot finish the task. Drafts, vibration or a moving sample usually cause this.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the draft shield doors.
  2. Move the balance to a solid, vibration-free bench.
  3. Keep it away from air conditioning, fans and doors.
  4. Select a less stable environment setting in the menu.

Possible causes

  • Drafts or air conditioning
  • Vibration on the bench
  • Moving or evaporating sample
  • Environment setting too sensitive

Do you need service?

If the reading still drifts with an empty pan in calm conditions, the weighing cell must be checked by a technician.

Models showing this code: ML104, ML204, ML303, ML3002

Source: Mettler Toledo official user manual · Open manufacturer document