EEPROM memory error
An error was detected in the EEPROM memory where the balance stores its adjustment and device data. Because the data in this memory is not reliable, the weighing results cannot be trusted.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Switch the balance off, unplug it, and switch it back on after 30 seconds. If the message appears again, contact service.
- Do not open the housing. The internal parts cannot be repaired by the user.
- Note the error code and serial number and report them to technical service.
Possible causes
- EEPROM memory failure
- Excessive fluctuation or sudden interference in the mains voltage
Do you need service?
An EEPROM memory error is related to the electronics board and cannot be fixed by the user. technical service is required.
Models showing this code: ME, NewClassic ME
Source: Mettler Toledo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.