Message on display Taring or zeroing was interrupted Mettler Toledo XS Analytical

Taring or zeroing was ınterrupted

The balance could not get a stable value within the settling time, so it aborted the operation. Drafts or vibration are the usual cause.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the draft shield doors.
  2. Check the location for drafts and vibration.
  3. Press OK.
  4. Repeat the taring or zeroing.

Possible causes

  • Draft shield doors open
  • Air drafts
  • Vibration on the bench

Do you need service?

If the operation keeps aborting in calm conditions, the weighing cell must be checked by a technician.

Source: Mettler Toledo official user manual