Code on display Cannot start adjustment Mettler Toledo XPR Precision Balances and Comparators

Calibration cannot be started

Adjustment (calibration) cannot be started because the initial zero could not be reached when the balance was switched on. The operation is blocked because an adjustment made without a verified zero point would be incorrect.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and check that it is installed correctly and does not touch anything.
  2. Unplug the power cable, wait a few seconds and plug it in again.
  3. Restart the adjustment after the balance has been switched on.

Possible causes

  • A load was on the weighing pan at switch-on
  • The weighing pan is installed incorrectly or is touching something
  • Vibration or air drafts during switch-on

Do you need service?

If the initial zero still cannot be taken after a restart, a service check of the weighing cell is required.

Models showing this code: XPR Precision