Calibration cannot be started
Adjustment (calibration) cannot be started because the initial zero could not be reached when the balance was switched on. The operation is blocked because an adjustment made without a verified zero point would be incorrect.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan and check that it is installed correctly and does not touch anything.
- Unplug the power cable, wait a few seconds and plug it in again.
- Restart the adjustment after the balance has been switched on.
Possible causes
- A load was on the weighing pan at switch-on
- The weighing pan is installed incorrectly or is touching something
- Vibration or air drafts during switch-on
Do you need service?
If the initial zero still cannot be taken after a restart, a service check of the weighing cell is required.
Models showing this code: XPR Precision
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.