Entered value out of range
In the statistics application, the sample weight placed was not recorded because it is outside the accepted range. The next sample must be within approximately 70%–130% of the current mean.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Empty the weighing pan.
- Place a new sample close to the current mean (within approximately 70%–130%).
- Start a new statistics series for samples of different weights.
Possible causes
- The sample differs too much from the current mean
- The wrong sample, or more than one sample, was placed
Do you need service?
The weighing is accepted when a sample within the proper range is placed. No service is required.
Models showing this code: MS-S
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.