Code on display NO STABILITY Mettler Toledo MS-L

Unable to obtain a stable weighing value

During taring, zeroing or adjustment (calibration), the balance waits for the reading to settle. Because the value keeps fluctuating, the operation cannot be completed. According to the manual, the solution is to improve the ambient conditions, or if that does not help, to change the environment setting.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the draft shield doors and block drafts from air conditioners, fans and windows near the balance.
  2. Move the balance to a solid, vibration-free weighing table and level it again using the bubble level.
  3. Make sure nothing (cable, draft shield part, sample container) touches the weighing pan.
  4. If conditions cannot be improved, set the environment (ENV) setting in the menu to a level suited to less stable environments and repeat the operation.

Possible causes

  • Vibration on the table or floor
  • Air drafts or an open draft shield door
  • Sample evaporation or static electricity
  • Environment setting not suited to the current conditions

Do you need service?

If the value still does not settle after the environment has been improved and the environment setting changed, the weighing cell must be checked by service.

Models showing this code: MS-L