Weighing cell data error
The data stored for the weighing cell (load cell) is invalid or incompatible with the device. Without this data, the balance cannot calculate weight correctly.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Switch the balance off, unplug it, and switch it back on after 30 seconds. If the message appears again, contact service.
- Do not open the housing. The internal parts cannot be repaired by the user.
- Note the error code and serial number and report them to technical service.
Possible causes
- Corruption of the weighing cell data
- Incompatible data after an electronics board or weighing cell replacement
Do you need service?
Reloading the weighing cell data requires service software, so technical service must be contacted.
Models showing this code: MS Semi-Micro
Source: Mettler Toledo official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.