Code on display BELOW INITIAL ZERO RANGE Mettler Toledo MS-S

Below ınitial zero range

When the balance took its initial zero at switch-on, the measured value was below the permitted range. According to the manual, the cause is the wrong weighing pan or a weighing pan that is not installed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Install the weighing pan and pan support and check that they are fully seated.
  2. Make sure the weighing pan in use is the original pan for this balance model.
  3. Switch the balance off and on again.

Possible causes

  • The weighing pan is not installed
  • A (lighter) weighing pan that does not belong to this model is used
  • The pan support is missing

Do you need service?

If the error persists although the correct weighing pan is installed, a service check is required for zero point drift.

Models showing this code: MS-S

Source: Mettler Toledo official user manual · Open manufacturer document