Gram precision scale error codes

5 series · 0 error codes

Barcelona-based Gram Precision (GRAM Group) is a Spanish manufacturer offering laboratory and analytical balances, moisture analyzers, and industrial and retail weighing solutions. The guide covers the error messages for the TGI/TG precision and analytical balances, FS analytical balance, TH moisture analyzer, EM industrial precision balance and RK counting scale.

ERR0 Zero range exceeded RK · When the zero key was pressed, the load on the weighing pan was above the zeroing limit (2% of full capacity). ERR1 Model setting error RK · The capacity and resolution (division) settings form an invalid combination. ERR2 Initial zero point exceeded RK · At startup the scale detected that the zero point is outside the permitted limit (30% of full capacity). ERR3 A/D resolution range exceeded RK · The signal from the load cell is outside the range the A/D converter can measure. ERR4 EEPROM checksum error RK · A checksum error was detected in the EEPROM memory that stores adjustment and calibration data. ERR5 Overload RK · The load on the weighing pan has exceeded the scale's maximum capacity. ERR6 Display range exceeded RK · The value to be displayed exceeds the display's digit capacity. ERR7 Accumulation capacity exceeded RK · The number of accumulated weighings or the total value exceeded the display range. ERR8 Limit value exceeds the capacity RK · The weight limit entered for checkweighing is greater than the scale's full capacity. ERR9 Tare range exceeded RK · The tare or preset tare value is outside the permitted range. ERR10 Incorrect calibration weight RK · The weight placed during calibration does not match the expected value. ERR01 Unstable weight after tare TGI / TG · The weight did not become stable after tare was taken. ERR02 Calibration did not start due to instability TGI / TG · Calibration could not be started because the balance is unstable. ERR03 Wrong calibration weight TGI / TG · The calibration weight is wrong or the balance is unstable during calibration. ERR04 Piece counting sample insufficient TGI / TG · The sample weight placed for piece counting is insufficient or unstable. ERR05 Unstable weight, could not print TGI / TG · The data could not be sent to the printer because the weight is not stable. ERR06 Instability in density mode TGI / TG · The weight could not stabilize in density determination mode. ERR07 Instability in percent mode TGI / TG · The weight could not stabilize in percent weighing mode. ERR08 Automatic calibration fault TGI / TG · An abnormality was detected in the internal automatic calibration system (motor/mechanism). ERR09 Instability in formulation mode TGI / TG · The weight could not stabilize in formulation mode. ERR10 Component out of tolerance TGI / TG · The component weight is out of tolerance in formulation mode. ERR F Flash memory damaged TGI / TG · The balance's flash memory is damaged. UNLOAD Empty the weighing pan warning TGI / TG · There is a load on the weighing pan or the pan is not seated correctly. CAL But Recalibration required TGI / TG · The balance reports that it needs to be recalibrated. ERR01 Unstable weight after tare TH · On the moisture analyzer, the weight did not stabilize after tare was taken. ERR02 Calibration could not be started TH · Calibration could not be started because the balance is unstable. ERR03 Incorrect calibration weight TH · The calibration weight is wrong or the instrument is unstable during calibration. Err-1 Switched on and off too often EM · The balance was switched on and off at very short intervals. Err-2 Balance not stable EM · The balance could not obtain a stable reading. P FAIL Power loss warning FS · The balance lost power without being switched off with the ON/OFF key. CAL-no Adjustment (calibration) error FS · The adjustment (calibration) procedure failed on the analytical balance.