Code on display CAL-no Gram Precision FS

Adjustment (calibration) error

The balance did not save the adjustment because it could not read the calibration weight at the expected value and stably. At 0.1 mg resolution, environmental conditions directly affect the result.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the draft shield doors and eliminate air currents.
  2. Wait for the balance to complete its 20 to 30 minute warm-up.
  3. Repeat the adjustment with a weight of the correct class for the model (for example 200 g for FS-220).

Possible causes

  • Wrong or dirty calibration weight
  • Air currents, vibration or temperature change
  • Warm-up time not observed

Do you need service?

If it persists when repeated with suitable conditions and weight, a service check is needed.

Models showing this code: FS-120, FS-220

Source: Gram Precision official user manual