Adjustment (calibration) error
The balance did not save the adjustment because it could not read the calibration weight at the expected value and stably. At 0.1 mg resolution, environmental conditions directly affect the result.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Close the draft shield doors and eliminate air currents.
- Wait for the balance to complete its 20 to 30 minute warm-up.
- Repeat the adjustment with a weight of the correct class for the model (for example 200 g for FS-220).
Possible causes
- Wrong or dirty calibration weight
- Air currents, vibration or temperature change
- Warm-up time not observed
Do you need service?
If it persists when repeated with suitable conditions and weight, a service check is needed.
Models showing this code: FS-120, FS-220
Source: Gram Precision official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.